Services Available |
---|
The Model 44212 (Air/Soil/Water) and 44212EC (Air) Temperature probes use thermistors to measure temperature. The probes contain a YSI 44212 thermilinear network. The 44212 and 44212EC probes are designed to be housed in the 41303-5A six plate Gill radiation shield or equivalent. The radiation shield is used to mount the 44212 and 44212EC probe and limit solar radiation loading. The 44212 temperature probe is also designed to be buried or submerged in water up to 200 feet. A ten foot lead length is standard on the 44212. Custom lead lengths are available up to 1000 feet.
Read MoreThe Model 44212 (Air/Soil/Water) and 44212EC (Air) Temperature probes use thermistors to measure temperature. The probes contain a YSI 44212 thermilinear network. The 44212 and 44212EC probes are designed to be housed in the 41303-5A six plate Gill radiation shield or equivalent. The radiation shield is used to mount the 44212 and 44212EC probe and limit solar radiation loading. The 44212B temperature probe is electrically identical, but is designed to be buried or submerged in water up to 200 feet. A ten foot lead length is standard on the 44212. Custom lead lengths are available up to 1000 feet.
YSI Thermilinear Network Specifications
-50 to +50°C
T1=2,000 Ω @ 25°C
T2=15,000 Ω @ 25°C
T3=45,000 Ω @ 25°C
55°C (131°F)
- ±0.1°C 44312
R1 = 23,100 Ω
R2 = 88,200 Ω
R3 = 38,000 Ω
±0.15°C @ -50°C, ±0.03°C @ +50°C
The 44212 should be housed in a 41303-5A 6-plate naturally-aspirated solar radiation shield. This radiation shield mounts to a tripod or tower mast, crossarm, or user-supplied pole.
CR200(X) Series | CR800/CR850 | CR1000 | CR3000 | CR5000 | CR7X | CR9000X |
CR500 | CR510 | CR10 | CR10X | 21X | CR23X | CR9000 |